Location: Products >> Emission >> Elemental Analyzers >> S-in-O >> Specifications
Sulfur-in-Oil Analyzers - SLFA Series Specifications
Model

SLFA-2800/2100

SLFA-20

SLFA-UV21

Range 0-10 weight % 0-5 weight % 0-1 weight %
Repeatability 5 ppm 15 ppm 1.6 ppm
Lower detection limit 5 ppm 20 ppm 30 ppb
Measuring principle Non-dispersive X-ray fluorescence (NDXRF) Non-dispersive X-ray fluorescence (NDXRF) Combustion ultraviolet fluorescence method
C/H ratio error ±50 ppm ±50 ppm N/A
Calibration curve Automatic and manual selection from five curves; automatic and manual selection of linear or parabolic format Automatic and manual selection from five curves; automatic and manual selection of linear or parabolic format Automatic and manual selection from five curves
Calibration points 2 to 20 calibration points per curve 2 to 20 calibration points per curve 2 to 20 calibration points per curve
Sample cell Disposable, sealed container Disposable, sealed container Disposable, sealed container
Sample volume 4 to 10 mL 4 to 10 mL 5-20 µL
Measurement time 10, 30, 100, 300 or 600 seconds, selectable 10, 30, 100, 300 or 600 seconds, selectable Approximately three minutes
Spectrum measurement Available Available N/A
Data output Built-in printer and RS-232C output Built-in printer and RS-232C output Built-in printer and RS-232C output
 