- Automated and batch titrations
- Controllable and programmable aliquot sizes and mixing speeds
- Sample preparation and experiments time-saver
- Eliminate technician to technician method variability
The
AutoTitration Injector is used for running automated and
batch titrations with precisely controllable
and programmable aliquot sizes and mixing
speeds.
The left syringe is used to inject aliquots of liquid into the sample to be scanned. The right syringe is used to mix the sample to be scanned, or to extract aliquots from the sample.
Data are shown as a three-dimensional plot, using the z-axis to represent total volume of titrant delivered
System Operation
The
F-3006 AutoTitrator is fully operated through the JY-SPEX® DataMaxT software.
There are two ways to alter the operation of the injector and syringes:
To adjust the basic injector parameters:
- Volume of the syringes
- How fast to aspirate the syringes
- Filling and emptying the left syringe
To modify a batch titration experiment:
- How many aliquots to add
- What volumes they ought to be
- Whether to mix the sample via the right syringe
Specifications
| Accuracy: | within ±1% |
| Precision: | within +0.2% |
| Resolution: | 0.1% of syringe volume |
| Vol. increment: | 0.1 to 100% of total syringe vol. |
| Speed: | 1 to 250 seconds per full syringe stroke |
| Syringe Drive Mechanism: | stepper motor driven high precision lead screw with encoder |
| Electrical Specifications: | 220 VAC at 0.7 A |
| Dimensions (cm): | 33.3 H × 15.6 W × 17.8 D |
| Communication interface: | RS-232 |
| Computer requirements: | 973005 from JY or equivalent computer with free RS-232 port |
| Software requirements: | .
Windows™, version 9# or NT4 (SP3 or higher) . DataMax™, version 2.2.10 |
| Instrument compatibility with: | .
FluoroLog-3® and FluoroMax-3® . FluoroLog-2® and FluoroMax-2® with appropriate sample chamber covers for tubing access |
| Certification: | CE |