Features
- Second generation device: more sensitivity ...
- For plates with up to 384 wells
- Compatible with FluoroMax and FluoroLog spectrofluorometers
- Full automatic
- Controlled by DataMax software
The MicroMax 384 is a microwell-plate reader able to accept plates with up to 384 wells, and connect to FluoroMax or FluoroLog spectrofluorometers.
MicroMax 384's high speed allows it to scan completely a 96-microwell plate in less than one minute. By moving the microwell plate through stationary optics, the MicroMax 384 insures high sensitivity, excellent accuracy, and high reproducibility.
Light from the excitation and emission monochromators is carried via a fiber-optic bundle to and from the MicroMax 384, thus the user may scan with the main spectrofluorometer and select any excitation and emission wavelength pair for intensity measurements.
All control of the MicroMax 384 is automated through DataMax software; custom selection of microwells on the plate is possible through the software.
Principle

Software Operation
DataMax™ applications specifically for the MicroMax 384.
The MicroMax 384 uses four applications within DataMax:
- Visual Instrument Setup
- Real Time Display
- Run Experiment
- Constant Wavelength Analysis
Specifications
- Wavelength Range: full range of spectrofluorometer's: 250-850 nm (other wavelength range are available)
- Microplate Acceptance: 384-wells (24 X 16) and 96-well (12 x 8). For non standard sizes, contact
- Speed: scans 96 wells in less than 1 minute
- Typical Sensitivity:
- 25 pM fluorescein with FluoroLog-322
- 10 nM fluorescein with FluoroMax-3
- Fiber Optics: Bifurcated-slit-end bundles. 1m length. Randomised distribution.
- Dimensions (cm): 17.8 H X 30.5 W X 20.3 D
- Electrical Specifications: 220 VAC at 0.25 A - 50 or 60 Hz single phase
- Communication Interface: RS-232
- Hardware Requirements: Fiber-optic adapter
- Computer Requirements: 973005 from HORIBA Jobin Yvon or equivalent PC computer with free RS-232 port
- Software Requirements:
- Windows, version 9# or NT4 (SP3 or higher)
- DataMax, version 2.2.10
- Instrument Compatibility with:
- FluoroLog-3 and FluoroMax-3
- FluoroLog-2 and FluoroMax-2 with appropriate fiber-optic adapter
- Certification: CE