Location: Products >> Gratings, OEM & VUV >> Vacuum UV >> Monochromators & Spectrographs >> Toroidal Grating Spectrograph
Vacuum UV - Toroidal Grating Spectrograph (TGS)
TGS Optical Layout

TGS Optical Layout

  • High luminosity : only one optical piece (aberration corrected toroidal holographic grating).
  • Medium resolution.
  • Grating fixed position.
  • Flat focal field to accomodate multichannel detector.
  • Spectral range : from 170 nm to 10 nm.
  • Exit arm length : 685 or 306 mm.
  • Deviation angle : 140°.

Standard Configuration

TGS 330

  • Deviation angle 140°
  • La = 292 mm
  • Lb = 330 mm
  • Grating blank dimensions = 12 x 34 x 10 mm3
  • Flat field on 40 mm
  • Spectral Resolution 5 to 1 Å measured with MCP + CCD

Spectral range

Grooves/mm

95-320Å (130-39eV)
2105
100-1100Å (124-11eV)
450
155-1700Å (80-7eV)
290
 