The
H10 Monochromator is a small, versatile instrument and has
several applications. You can use it as follows:
- As a Variable Filter the H10 is very useful in applications for selecting a range of wavelengths from a continuous source. With its metal drop-in slits, it can be used with focused arc lamps up to 1Kw.
- Low resolution absorption transmission and fluorescence applications
The H10 features HIGH SPECTRAL PURITY. High stray light level rejection of 10-5 at 8 nm from the 632.8 nm laser line.
We offer a wide range of accessories for optical configurations including light sources, detectors, optical benches, motors, recorders and computer interfaces.
Features
Small size: just 10 cm3
Featherweight: 1Kg
In spite of its small size, the H10 monochromator offers
performance equal to that of bulkier instruments.
Optical Configuration
The H10 is built around an aberration-corrected concave holographic diffraction grating. This element plays a triple role: collimator, dispersing and focusing system.
Resolution
Typically 1 nm at 500 nm
High Luminosity
This is assured by its f/3.5 aperture
Slits
Manually interchangeable
fixed width slits, with continuous height adjustment.
Custom OEM version available.
For applications where resolution is required optional fixed slits are available. Each slit is equipped with a height limiter. Maximum height: 8 mm, minimum height: 2 mm.
Specifications
| Focal length | 100mm |
| Aperture Ratio | f/3.5 |
| Grating | 30 x 30 mm, type IV, aberration corrected, concave, holographic |
| Bandpass | 4 nm with 0.5 mm slits and 1200 g/mm gratings |
| Slits | Fixed width, interchangeable. 3 pairs supplied with instrument: 0.5, 1 and 2 mm. |
| Stray Light | 10-5 at 8 nm from 632.8 nm laser line |
| Resolution | 1 nm with 0.1 mm slits and 1200 g/mm grating |
| Wavelength counter | Mechanical, direct reading in nm, graduated in 0.2 nm intervals |
| Reproducibility | ± 0.25 nm with 1200 g/mm grating |
| Dimensions | 128 x 114 x 84 mm (l x w x h) |
| Weight | 1 Kg |